IEC 61223-3-5 Ed. 1.0 b:2004, Evaluation and routine testing in medical imaging departments - Part 3-5: Acceptance tests - Imaging performance of computed tomography X-ray equipment Buy on Amazon
Facebook LinkedIn

IEC 61223-3-5 Ed. 1.0 b:2004, Evaluation and routine testing in medical imaging departments - Part 3-5: Acceptance tests - Imaging performance of computed tomography X-ray equipment

Price not available for India

You can still browse on Amazon. Try another country above.

Book Details
Author(s) IEC TC/SC 62B
ISBN / ASIN B000Y2LSB4
ISBN-13 978B000Y2LSB0
Marketplace India 🇮🇳
Description
This standard - defines the essential parameters which describe the performance of the CT scanners with regard to image quality, patient dose and positioning; - defines the methods of testing the essential parameters; - evaluates compliance with the tolerances of the parameters specified by the accompanying documents. These methods rely mainly on non-invasive measurements, using appropriate test equipment, performed during the installation or after it has been completed. Signed statements covering steps in the installation procedure may be used as part of the acceptance test report. This part of IEC 61223 is intended to assist in performing the acceptance tests on a CT scanner. The aim is to verify compliance of the installation with specifications affecting the image quality, patient dose and positioning.
Donate to EbookNetworking
No Prev
No Next